Study Notes: TG 142 - Quality assurance of Medical accelerators

July 10, 2026  |  Saikat Roy
Linear Accelerator QA Tolerances - Exam Dashboard

RTonc.site the Exam Preparation: LINAC QA

Task Group 142 report: Quality assurance of medical accelerators 2009

TG-142 Ref
Most Common and Important
Parameter Tolerance / Requirement
Laser±2 mm
ODI±2 mm
Couch movement / Jaw position±2 mm
Couch / Gantry / Collimator angle±1°
MLC leaf position±1 mm
Light vs radiation field±2 mm
Daily / Monthly / Annual Output±3% / ±2% / ±1%
Imaging coincidence±2 mm
SRS isocenter vs Conventional≤1 mm vs ≤2 mm diameter
Daily QA
Parameter 3DCRT IMRT/VMAT SRS/SBRT
Output constancy±3%±3%±2%
Laser localization±2 mm±2 mm±1 mm
ODI±2 mm±2 mm±1 mm
Imaging & treatment coincidence±2 mm±2 mm±1 mm
Door interlock / Beam-onFunctional
Audiovisual monitorFunctional
Monthly QA
Parameter Tolerance / Requirement
Output constancy (X-ray / Electron)±2%
Beam flatness & symmetry±2%
Laser alignment / ODI / Cross-hair±2 mm
Light vs radiation field / Jaw position±2 mm
Couch vertical/longitudinal/lateral±2 mm
Couch / Gantry / Collimator rotation±1°
Mechanical QA
Parameter Tolerance / Requirement
Laser / ODI / Cross-hair±2 mm
Light & Radiation field size / Coincidence±2 mm
Jaw calibration / Couch motion±2 mm
Gantry / Collimator / Couch indicators±1°
MLC QA
Parameter Tolerance / Requirement
Leaf position accuracy & reproducibility±1 mm
Dynamic leaf positioning / Picket Fence±1 mm
Leaf transmissionBaseline consistency
Leaf speedWithin baseline
SRS/SBRT Specialized QA
Parameter Tolerance / Requirement
Winston–Lutz Test≤1 mm
Radiation isocenter≤1 mm
Laser / ODI / Couch position±1 mm
Imaging coincidence / MLC positioning±1 mm
Output constancy±2%
Imaging Systems QA
Parameter Tolerance / Requirement
kV imaging coincidence±2 mm
MV imaging coincidence±2 mm
CBCT coincidence±2 mm
Imaging vs treatment isocenter±2 mm
Annual QA
Parameter Tolerance / Requirement
Absolute dose calibration±1%
Photon / Electron output±1%
Beam quality / Flatness / Symmetry±1%
Radiation isocenter≤2 mm diameter
Electron Beam QA
Parameter Tolerance / Requirement
Output constancy / Symmetry±2%
Beam energy (R50/practical range)±2 mm
Flatness±3%
Applicator positioning±2 mm
Safety Systems Interlocks
Door Interlock
Emergency OFF
Beam Interrupt
Collision Interlock
Audiovisual Mon.
Rad Monitor
R&V Comm.
MUST FUNCTION